Compliance Test of X-Ray Beam Quality Using HVL Parameter on Digital Mammography
Main Article Content
Abstract
The compliance test of X-ray beam quality using HVL parameter on digital mammography was determined at a hospital in Makassar in 2019-2021. This measurement uses a ruler, X-ray multimeter, and aluminum filter. The exposure process is carried out at certain kVp and mAs conditions after placing the detector on the examination table. The HVL test to determine the X-ray beam quality is carried out by placing the detector 4 cm from the edge of the chest wall on the patient's table in an upright position facing the tube. The resulting HVL values are 0.37, 0.37, and 0.38. This study resulted in a p-value of 0.000 (p < 0.05). Statistically, there is a significant difference in HVL values in 2019-2021. The resulting HVL value is by the tolerance limit recommended by the IAEA, which is 0.31 ≤ HVL ≤ 0.58. These values indicate that mammography works well, and it is safe for the patient.
Article Details
References
Fitriani, M. Yunus, M. Fachrul, A. I. W. S. Ramadani, “Determination of Suitability Test on Digital Mammography Using Voltage Accuracy Parameter,” vol. 02, no. 01, pp. 2774–2777, 2021.
M. Yunus and A. Arifin, “Karakterisasi Sensor Kekentalan Oli Berbasis Serat Optik Plastik Menggunakan Metode Back Scattering,” Positron, vol. 8, no. 1, p. 31, 2018.
M. Yunus, A. Arifin, N. Fuadi, and A. M. Syafar, “Sensitivity Enhancing of Oil Viscosity Sensor Based on Optical Fiber by Using Bending Angle Method,” INTEK J. Penelit., vol. 8, no. 1, p. 43, 2021.
Y. M. Y. Abdallah, A. O. Alamoudi, and S. N. A. Elgak, “International Journal of Advanced and Applied Sciences Appraisal of mean glandular dose (MGD) during mammography examinations,” Int. J. Adv. Appl. Sci., vol. 4, no. 2, pp. 49–51, 2017.
Fitriani, B. Abdullah, D. Tahir, “Studi Uji Kesesuaian Selisih Lapangan Kolimasi pada Pesawat Mammografi Analog dan Digital,” Flux, vol. 16, no. 2, p. 5, 2019.
A. Amir, M. Yunus, and M. F. Latief, A. I. W. S. Ramadani, “X-Ray Technology to Evaluate Mortality of Covid-19 Patients in Comorbid Diabetes Mellitus and Stroke,” vol. 02, no. 02, pp. 2774–2777, 2021.
Fitriani, B. Abdullah, and D. Tahir, “Measurement of X-Ray Beam Quality and Patient Dose Information from Analog Mammography,” IOP Conf. Ser. Mater. Sci. Eng., vol. 546, no. 3, 2019.
C. P. Nguyen and E. M. M. Adang, “Cost-Effectiveness of Breast Cancer Screening Using Mammography in Vietnamese Women,” PLoS One, vol. 13, no. 3, pp. 1–17, 2018.
M. Zeeshan, B. Salam, Q. S. B. Khalid, S. Alam, and R. Sayani, “Diagnostic Accuracy of Digital Mammography in the Detection of Breast Cancer,” Cureus, vol. 10, no. 4, 2018.
J. Dlama Zira and C. Chukwuemeka Nzotta, “Radiation Doses for Mammography and its Relationship with Anthropo-Technical Parameters,” Int. J. Radiol. Radiat. Ther., vol. 5, no. 1, pp. 14–18, 2018.
H. Al-Naemi et al., “Evaluation of Mean Glandular Dose from Digital Mammography Exams at Qatar and Compared with International Guidelines Levels,” Br. J. Med. Med. Res., vol. 14, no. 2, pp. 1–9, 2016.
J. Arif, “Analisa Uji Kesesuaian Pesawat Sinar-X Radiografi,” Jurnal.Umj.Ac.Id, pp. 0–7, 2019.
A. Mutmainna, S. D. Astuty, S. Dewang, and Mulyadin, “Uji Kesesuaian Standar Nilai HVL Filter Aluminium pada Pesawat Sinar-X Mammografi : Studi Kasus di Ruang Instalasi Radiologi RS. Siloam Makassar,” Berk. Fis., vol. 23, no. 1, pp. 17–25, 2020.
H. D. Yunitasari and E. Setiawati, “Evaluasi Metode Penentuan Half Value Layer (HVL) Menggunakan Multi Purpose Detector (Mpd) Barracuda Pada Pesawat Sinar-X Mobile,” Youngster Phys. J., vol. 3, no. 2, pp. 113–118, 2014.
C. Pwamang, E. Sosu, and C. Schandorf, “Assessment of Dose to Glandular Tissue of Patients Undergoing Mammography Examinations,” J. Radiol. Radiat. Ther., 2016.
S. Hyperastuty, Agoes Santika, Yanuar Mukhammad, “Analisis Uji Kesesuaian Pesawat Sinar X Radiografi Mobile Merk Drgem Topaz-40d Menggunakan X-Ray Multimeter PIRANHA,” J. Heal. Sci., vol. 29, no. 3, pp. 419–429, 2563.
N. P. P. A. Yulianti, G. N. Sutapa, and K. Y. Astina, “Uji Kesesuaian Kualitas Berkas Sinar-X Menggunakan Half Value Layer (HVL) Filter Aluminium Pada Pesawat Sinar-X Stationary di Akademi Teknik Radiodiagnostik dan Radioterapi Bali,” Bul. Fis., vol. 24, no. 1, p. 27, 2021.
International Atomic Energy Agency, “Quality Assurance Programme for Screen Film Mammography,” IAEA Hum. Heal. Ser., no. 2, 2009.